
- Transmittance Definition: Transmittance is the ratio of light intensity passing through a material to the light intensity hitting the material’s surface.
- Transmittance Formula: It is calculated by dividing the intensity of the light that passes through the object by the intensity of the incident light.
- Radiant Flux Formula: Another way to calculate transmittance is by dividing the transmitted radiant flux by the received radiant flux.
- Absorbance Relationship: According to Beer-Lambert law, absorbance is equal to two minus the logarithm base ten of the percentage transmittance.
- Applications of Transmittance: Includes measuring chemical concentrations, water clarity, and window tinting.
What is Transmittance?
The transmittance of a surface or material is defined as the part of the light that moves through the other side of the surface. When light passes through any surface or material, it can be transmitted, reflected, or absorbed. Transmittance and reflectance are closely related concepts.
Transmittance is defined as the ratio of the intensity of incident light (I0) to the intensity of light that passes through the object (I). It is denoted as T.
![]()
![]()

As shown in the above figure, the I0 is the intensity of incident light. This light passes through the block of glass or any other material. The I is the intensity of light that can pass through the material.
Since transmittance is a ratio of intensities, it has no unit.
Let’s understand the transmittance by an example.
If you pass the light from a semi-transparent block of glass. Let say 30% of light is reflected from the surface of the glass. Remaining 70% of the light will try to pass the block of glass. But, some amount of light will be absorbed by the molecules of light. Let say it is 35% of light. Now the remaining 35% light will pass from the other side of the block. It means that the glass block has a transmittance of 35%.
Transmittance Formula
Transmittance is also defined in terms of radiant flux. The ratio of radiant flux transmitted by surface (фt) to the radiant flux received by that surface (фi) is known as the Transmittance (T). So, the formula of transmittance is;
![]()
Transmittance to Absorbance
The amount of radiation absorbed can be measured by the transmittance. The relation between transmittance (T) and absorbance (A) is given by the Beer-Lambert law (Beer’s law).
According to Beer’s Law, the absorbance is formulated as:

If light passes through an object without any absorption, 100% of the light will pass through, resulting in a transmittance of 100%.
From the equation of Beer’s law, we can calculate the absorbance and it is zero.
Now assume the opposite condition – the light cannot pass the object. In this condition, the transmittance is zero and the absorbance is infinite.
Absorbance vs Transmittance
Absorbance and transmittance both terms are opposite to each other. The difference between these two terms is summarized in the table below.
| Transmittance | Absorbance | |
| Definition | Transmittance is a ratio of the incident intensity of light (I0) to the amount of intensity passes through the object (I). | Absorbance is defined as the amount of light absorbed by the molecules of the object. |
| Equation | | |
| How the value change as the concentration is increased | Transmittance decreases exponentially. | Absorbance increase linearly. |
| Graph | ![]() | ![]() |
| Range | Values range from 0 to 1 and the percentage transmittance range from 0% to 100%. | Absorbance takes values from 0 upwards. |
What is Percentage Transmittance
Transmittance measures the amount of light that can pass through the material. Percentage transmittance is defined as the percentage of light that can be transmitted from the other side of the surface.
The equation of percentage transmittance (%T) is as
![]()
What Does a High Percent Transmittance Mean
The value of percentage transmittance provides information about surface or material used in various applications.
If the value of percentage transmittance is high, it means that the surface will allow more lights to pass. Similarly, if the value of percentage transmittance is low, it means that the surface absorbs the more amount of light.
For a perfectly transparent material, 100% of the light will pass through, giving it a transmittance of 100%.
For a completely opaque material, the light cannot pass through the material. Therefore, the transmittance is 0%.
So, high the value of percentage transmittance gives higher the transparency of the material.
Why is Absorbance the Preferred Unit Over Transmittance?
The absorbance and transmittance are related to each other. The old IR-spectra can only be found in Transmittance mode. So, it only measures the Transmittance. But, after the use of microprocessors, it automatically converts transmittance into absorbance.
The use of absorbance and transmittance is a personal preference.
Absorbance is a logarithmic scale of Transmittance. The absorbance is enabled to detect weak signals next to pass signals.
And also, the absorbance is proportional to analytic concentration whereas the transmittance is not proportional to analytic concentration.
What is the Use of Transmittance (Applications)
The measurement of transmittance has many applications, including:
- Measuring concentrations of chemicals in solutions
- Water clarity
- Grade of syrup
- Testing window tint films and glass clarity
- Atmospheric haze





