- De Sauty Bridge Definition: De Sauty Bridge is defined as a method for comparing the values of two capacitors by balancing a bridge circuit.
- Working Principle: It operates by adjusting resistances to balance the voltage drops across capacitors, ensuring accurate comparison.
- Balance Achievement: To achieve balance, the values of specific resistors are adjusted without altering other elements in the bridge.
- Advantages and Limitations: The De Sauty Bridge is simple and easy to use but provides inaccurate results for capacitors with dielectric losses.
- Grover’s Modification: Adding resistances to the bridge can account for dielectric losses, improving accuracy for imperfect capacitors.

De Sauty Bridge provides a reliable way to compare two capacitor values when dielectric losses are ignored. The circuit of De Sauty’s Bridge is shown below.
A battery is connected between terminals 1 and 4. Arm 1-2 contains an unknown capacitor (c1) carrying current (i1). Arm 2-4 has a pure resistor (non-inductive), arm 3-4 also has a pure resistor, and arm 4-1 contains a known standard capacitor.
Let us derive the expression for capacitor c1 in terms of standard capacitor and resistors.
At balance condition we have,
It implies that the value of capacitor is given by the expression
In order to obtain the balance point we must adjust the values of either r3 or r4 without disturbing any other element of the bridge. This is the most efficient method of comparing the two values of capacitor if all the dielectric losses are neglected from the circuit.

Now let us draw and study the phasor diagram of this bridge. Phasor diagram of De Sauty bridge is shown below:
Let us mark the current drop across unknown capacitor as e1, voltage drop across the resistor r3 be e3, voltage drop across arm 3-4 be e4 and voltage drop across arm 4-1 be e2. At balance condition the current flows through 2-4 path will be zero and also voltage drops e1 and e3 be equal to voltage drops e2 and e4 respectively.


In order to draw the phasor diagram we have taken e3 (or e4) reference axis, e1 and e2 are shown at right angle to e1 (or e2). Why they are at right angle to each other? Answer to this question is very simple as capacitor is connected there, therefore phase difference angle obtained is 90o.
Now instead of some advantages like bridge is quite simple and provides easy calculations, there are some disadvantages of this bridge because this bridge give inaccurate results for imperfect capacitor (here imperfect means capacitors which not free from dielectric losses). Hence we can use this bridge only for comparing perfect capacitors.
Here we interested in modify the De Sauty’s bridge, we want to have such a kind of bridge that will gives us accurate results for imperfect capacitors also. This modification is done by Grover. The modified circuit diagram is shown below:
Here Grover has introduced electrical resistances r1 and r2 as shown in above on arms 1-2 and 4-1 respectively, in order to include the dielectric losses. Also he has connected resistances R1 and R2 respectively in the arms 1-2 and 4-1. Let us derive the expression capacitor c1 whose value is unknown to us. Again we connected standard capacitor on the same arm 1-4 as we have done in De Sauty’s bridge. At balance point on equating the voltage drops we have:
On solving above equation we get:
This the required equation.
By making the phasor diagram we can calculate dissipation factor. Phasor diagram for the above circuit is shown below
Let us mark δ1 and δ2 be phase angles of the capacitors c1 and c2 capacitors respectively. From the phasor diagram we have tan(δ1) = dissipation factor = ωc1r1 and similarly we have tan(δ2) = ωc2r2.
From equation (1) we have
on multiplying ω both sides we have

Therefore the final expression for the dissipation factor is written as
Hence if dissipation factor for one capacitor is known. However this method is gives quite inaccurate results for dissipation factor.