
- Notch Filter Defined: A notch filter (also known as a bandstop filter or reject filter) is defined as a device that blocks certain frequencies while allowing others to pass.
- Circuit Overview: The notch filter circuit combines low-pass and high-pass filters in a configuration that blocks specific frequency bands.
- Design Principles: Designing a notch filter involves selecting appropriate values for resistors, capacitors, and inductors to achieve the desired frequency rejection.
- Transfer Function Explained: The transfer function of a notch filter describes how it attenuates certain frequencies while maintaining others, crucial for understanding filter behavior.
- Practical Applications: Notch filters are essential in electronics to remove interference, such as the 50/60 Hz noise in medical ECG readings, enhancing signal clarity.
What is a Notch Filter (Bandstop Filter)?
A notch filter (also known as a bandstop filter or reject filter) is defined as a device that blocks specific frequency while allowing those outside the defined range to pass. Notch filters eliminate transmission of a narrow band of frequencies and allow transmission of all the frequencies above and below this band. As it eliminates frequencies hence, it is also called a band elimination filter.
A notch filter is essentially a band stop filter with a narrow stopband and two passbands. As in the band-pass case, a band-reject filter can be either wideband or narrow-band.
If the filter blocks a wide range of frequencies, it’s called a band-reject filter. If it blocks a narrow range, it’s known as a notch filter. The characteristics of a band-stop filter are exactly the reverse of the bandpass filter. Hence, the notch filter is a complement of the bandpass filter.
For example, if a Notch Filter has a stopband frequency from 100 MHz to 200 MHz, then it will pass all the signals from DC to frequency of 100 MHz and above 200 MHz, it will only reject frequency between 100 MHz to 200 MHz.
Thus, the function of a Notch Filter is to passing all those frequencies from zero (DC) up to lower cut-off frequency(fL) and above higher cut-off frequency(fH), and reject all those frequencies that lie in the bandwidth region i.e., BW= fH-fL.
Detection and filtering of narrowband signals in the presence of noise are the important applications of signal processing techniques. In many applications, it is necessary to remove the narrowband signal without changing band energy. This can be achieved by passing the signals through a notch filter.
Notch Filter Circuit
The notch filter is a combination of low pass and high pass filters similar to the bandpass filter design, but the difference is that we connect both the filters in parallel connection instead of cascading connection. The circuit diagram of the notch filter is shown in the below figure.

The upper section of the notch filter circuit includes a passive RC low pass filter, featuring two resistors (R1 and R2) and a capacitor (C1) in a ‘T’ configuration. This filter will allow the signals having frequencies lower than the higher cut-off frequency (fH).
The lower section features a passive RC high-pass filter, consisting of two capacitors (C2 and C3) and a resistor (R3), also arranged in a ‘T’ configuration. This filter will allow the signals having frequencies higher than the lower cut-off frequency (fL). This combination of both ‘T’ configuration is commonly known as a ‘Twin-T’ filter.
The typical configuration and a frequency response of a notch filter is shown in the figure below.


Notch Filter Types
There are many types of notch filter circuits are designed. Let’s explain the major types of filter circuits in detail.
Active Notch Filter
The active notch filter is a parallel combination of low pass filter and high pass filter with the op-amp as an amplifying component is shown in the figure below.

The circuit diagram of the Active Notch Filter is divided into three portions. The upper portion is an active low pass filter (LPF) which is in parallel to an active high-pass filter (HPF). The third and last portion of the circuit is the amplifying part in which op-amp is used for amplification. The final op-amp stage is a summing amplifier or a Op-Amp adder which acts as a summer. The circuit diagram of an active notch filter is shown in the figure below.
Here, the high-frequency cut-off of the low-pass filter is
![]()
while the low-frequency cut-off of the high-pass filter is
![]()
Passive Notch Filter
The passive filter comprises only passive components such as resistors, inductors, and capacitors and it does not use any active components such as op-amp for amplification. So, the amplification part is absent in a passive notch filter.
The passive notch filter is a combination of a passive low pass filter and a passive high pass filter. The circuit diagram of the passive notch filter is shown in the figure below.

The upper portion of the passive notch filter circuit is the passive low pass filter in parallel with the passive high pass filter.
Optical Notch Filter
An optical notch filter selectively blocks specific wavelengths while transmitting all others, crucial for various optical applications. It can be designed for various rejection levels, which is typically in terms of optical density.
In an optical system, specific wavelengths of light are selectively redirected by a diffraction grating or a dispersive prism. In the case of transmission gratings and prisms, polychromatic light that passes through the object will be redirected to wavelength, and then desired wavelengths can be achieved by the optical notch filter.
In the case, when using optics with real materials, the light will be attenuated at various wavelengths and it is filtered by interference with the medium through which the light passed. An optical notch filter passes wavelengths of a light that are unaltered or minimally attenuated.
Optical notch filters are used in laser-based Raman spectroscopy, laser-based fluorescence, and other biomedical and life science applications.
RLC Notch Filter
As the name suggests RLC, this notch filter or bandstop filter contains passive elements resistor, inductor, and capacitor. Therefore, it is also known as a passive notch filter. The circuit diagram of the RLC notch filter is shown in the below figure.

The shunt element of an RLC notch filter is a series combination of L and C. The output is taken across this series combination of inductor and capacitor. The equivalent circuit and characteristics of the RLC notch filter are shown in the figure below.

Here, the equivalent impedance of a series combination of the L and C is given by,
![]()
Now, the net impedance of the circuit is given by,
![]()
Now, applying KVL (Kirchhoff’s Voltage Law) in loop a-b-c-d to the equivalent circuit of RLC notch filter, we get,

Now, at the series resonance condition, the resonant frequency is given by
![]()
And the Q-factor is given by,
![]()
Where,
= Bandwidth = ![]()
while,
is the lowest cut-off frequency, and
is the highest cut-off frequency. The characteristics of the notch filter is shown in the figure below.

RLC type Notch Filter can also be made by utilizing the circuit that is shown in the figure below.

RF Notch Filter
A narrow notch filter is an RF notch filter that can be used to reject the carrier frequency during the measurement of non-linearities of power amplifiers.
An LC circuit notch filter is used to reject a specific interfering frequency in the radio frequency domain. The LC circuit notch filter circuit is used with radio receivers that are close to the transmitter that it swamps all other signals. The wave trap is used to reject the signal from the transmitter.
FM broadcast signals are very strong and it can prevent an SDR (Software-defined Radio) from processing weak signals. FM notch filters are very useful for SDR applications. Hence, The FM signals first pass through the notch filter, and then it is given to the SDR USB.

Butterworth Notch Filter
A Butterworth notch filter is a special type of signal processing filter designed to have a frequency response as flat as possible as in the narrow stopband.
An ideal Notch filter rejects some range of frequencies without distortion and passes all other frequencies. But in some applications, a simple notch filter is not accurate and reliable because of the low signal to noise ratio. In those cases, Butterworth’s notch filter is used to increase accuracy and reliability.
For example, In ECG (electrocardiogram) the noises that commonly disturb are power line interference, instrumentation noise, external magnetic field interference, noise due to random body movements, and respiration movements. These noises can be characterized based on their frequency content.
The bandwidth of the noise overlaps that of the desired signals so that ordinary filtering technique cannot sufficiently enhance the signal to noise ratio. Therefore, it is necessary to reduce these noises in the ECG signal to improve accuracy and reliability. The 4th order Butterworth notch filters with 3 dB stopband are used to reduce 50 Hz power line noise interference from ECG signals.
Notch Filter Design Example
An RLC type Notch Filter is to work with cut-off frequencies 23 kHz and 25 kHz. Assume Inductance L=45 mH, Design the RLC type Notch Filter.
Given data: fL=23 kHz, fH=25 kHz, L=45 mH = 0.045 H
- Bandwidth (BW) =

- Resonant frequency

- Now, at resonance condition resonant frequency is given by,

- Quality factor (Q) =

- Now,

Hence, the Notch filter configurations are
R=565.2 ohms , L=45 mH , C=977 pF
Thus, the design of the notch filter from the above configurations is shown in the below figure.

Notch Filter Transfer Function
The transfer function of the Notch Filter is given by

Where,
is the zero-circular frequency
is the pole-circular frequency
Q is the Quality factor. It is the selectivity of the filter.
where, BW is the bandwidth of the filter.
There are three cases in the characteristics of the notch filter. i.e., a standard notch, a low-pass notch, and a high-pass notch. The relationship between the zero-circular frequency
and the pole-circular frequency
determines the characteristics of the notch filter.
- If the pole-circular frequency is equal to the zero-circular frequency, i.e.,
, then the filter is the standard notch type. - If the pole-circular frequency is less than the zero-circular frequency, i.e.,
<
, then the filter is the low-pass notch type. - If the pole-circular frequency is greater than the zero-circular frequency, i.e.,
>
, then the filter is the high-pass notch type.
Notch Filter Applications
Some of the applications of the Notch Filter include:
- A Notch filter is generally used in communication systems, Instrumentation and control systems, and the biomedical field to eliminate 50/60 Hz power line interference.
- Notch filter or Bandstop filter is widely used in electronics and communications circuits to reject a band of unwanted frequencies and allowing transmission of other frequencies with minimum loss.
- Switching type of AC & DC motor drives, converters, and inverters cause sinusoidal disturbances at certain harmonics of the line frequency. The use of a notch filter eliminates such unwanted disturbances and enables accurate measurements.
- It is highly preferred in image and signal processing to reject unwanted frequencies i.e. noise.
- It is used in audio signal processing, for removing a specific range of unwanted frequencies i.e. noise or hum.
- It is used in telephone technology, DSL, and other internet services as a line noise reducer to reduce unwanted interference. Note that DSL is the Digital Subscriber Line that is used to transmit digital information over telephone lines.
- It is used in guitar amplifiers, instrument amplifiers, acoustic guitar, mandolin, bass instrument amplifier, and PA systems to reduce a specific humming sound that may produce after instruments are plugged. Note that PA (Public Address) systems are an electronic system comprising of microphones, amplifiers, loudspeakers, and other musical equipment.
- It is used in medical field applications i.e. in ECG (Electrocardiogram) measurements, to eliminate dc component.

